cawt monitor
Data-driven Design of Context-aware Monitors for Hazard Prediction in Artificial Pancreas Systems
Zhou, Xugui, Ahmed, Bulbul, Aylor, James H., Asare, Philip, Alemzadeh, Homa
Medical Cyber-physical Systems (MCPS) are vulnerable to accidental or malicious faults that can target their controllers and cause safety hazards and harm to patients. This paper proposes a combined model and data-driven approach for designing context-aware monitors that can detect early signs of hazards and mitigate them in MCPS. We present a framework for formal specification of unsafe system context using Signal Temporal Logic (STL) combined with an optimization method for patient-specific refinement of STL formulas based on real or simulated faulty data from the closed-loop system for the generation of monitor logic. We evaluate our approach in simulation using two state-of-the-art closed-loop Artificial Pancreas Systems (APS). The results show the context-aware monitor achieves up to 1.4 times increase in average hazard prediction accuracy (F1-score) over several baseline monitors, reduces false-positive and false-negative rates, and enables hazard mitigation with a 54% success rate while decreasing the average risk for patients.
- North America > Canada > Ontario > Toronto (0.14)
- North America > United States > Virginia > Albemarle County > Charlottesville (0.14)
- North America > United States > Florida > Alachua County > Gainesville (0.14)
- (5 more...)
- Information Technology > Security & Privacy (1.00)
- Health & Medicine > Therapeutic Area > Endocrinology > Diabetes (1.00)
- Health & Medicine > Pharmaceuticals & Biotechnology (1.00)
- Health & Medicine > Health Care Technology (1.00)